技術報告
ポリグラフ検査用呼吸トランスデューサの比較
小川 時洋松田 いづみ廣田 昭久高澤 則美
著者情報
ジャーナル フリー

2012 年 17 巻 2 号 p. 99-106

詳細
抄録
  The current study compared two types of transducers of respiratory movements used for psychophysiological recordings, evaluating their performance in capturing and reproducing variations in underlying respiratory movements. We examined a newly developed air-filled bellows-type transducer and compared its performance against the carbon-film transducer, which is currently used in polygraph examinations for criminal investigations in Japan. An experiment using a calibrator to control underlying movements revealed that the performance of the air-filled bellows-type transducer was superior to that of the carbon film-type transducer in capturing the morphology of respiratory waveforms as well as variations in their amplitude. A polygraph experiment (n=19) demonstrated that the air-filled bellows-type transducer produced differences with larger effect sizes between critical and non-critical items in respiratory measures compared with the carbon-film type transducer. These results indicated that the air-filled type transducer exhibited superior performance for the recording of respiratory movements.