Accuracy of One Minute Walking on Treadmill Exercise Test Using Automatic Measurements of Ankle-brachial Pressure Index
Kazuo Tsuyuki, Kenji Kohno, Kunio Ebine, Susumu Tamura, Yasuhiro Ohzeki, Toshifumi Murase, Kenji Ninomiya, Kenta Kumagai, Naoya Moriki, Itaru Yokouchi, Yoshiyuki Yazaki, Shinichi Watanabe
Author information
JOURNAL OPEN ACCESS

2012 Volume 52 Issue March Pages 175-179

Details
Abstract
Accuracy of one minute walking on treadmill exercise test (WT) using automatic measurements of ankle-brachial pressure index (ABI) is not known. Results of 5 and 1 minute WTs were compared with computed tomographic angiography findings in 57 patients who had suspected peripheral arterial occlusive disease (PAD). The sensitivity was 78.8% in both WTs. The specificity was 100% in 5 minute WT, and was 93.8% in 1 minute WT. Since the accuracy of 1 minute WT is almost equal with 5 minute WT, it is useful for a diagnosis of PAD.