Theoretical and Experimental Evaluation for Blood Pressure Drop and Oxygen Transfer Rate in Outside Blood Flow Membrane Oxygenator
Noriaki MatsudaMasayuki NakamuraKiyotaka SakaiKatsuyuki KuwanaKoichiro Tahara
著者情報
ジャーナル 認証あり

1999 年 32 巻 6 号 p. 752-759

詳細
抄録

An outside blood flow membrane oxygenator with an optimum blood flow pattern is developed by a trial and error method where large cost and long time are spent. We modeled the blood flow in the membrane oxygenator as the flow of low Reynolds number in a porous media and evaluated blood pressure drop and oxygen transfer rate using Darcy’s law. Governing equations are solved by the finite element method. Theoretical blood flow patterns agree well with visualized flow patterns obtained by X-ray computed tomography. Calculated values for blood pressure drop and oxygen transfer rate also agree with experimental data. Optimum structure parameters of the membrane oxygenator can hence be determined by simulation analysis.