Polarization Potentials Causing Pacemaker Oversensing
Andrzej OkreglickiToshio AkiyamaCeleste OcampoDennis Flynn
著者情報
ジャーナル フリー

1998 年 62 巻 11 号 p. 868-870

詳細
抄録

Pacemaker oversensing with inappropriate inhibition is commonly caused by sensing of either T-waves or polarization potentials, which are difficult to distinguish from one another. Decisive evidence, obtained from telemetered electrograms and subthreshold pacing, is presented to support sensing of the polarization potential alone and the exclusion of the T-wave as the cause of oversensing in a patient. (Jpn Circ J 1998; 62: 868 - 870)