2001 Volume 65 Issue 11 Pages 927-930
The study prospectively investigated the incidence, cause and efficient management of inappropriate discharge by the fourth generation implantable cardioverter-defibrillator (ICD) system in 45 patients (mean age, 57±16 years). During the follow-up period of 27±17 months, 18 patients (40%) experienced one or more inappropriate therapies: sinus and supraventricular tachycardia (15 patients) and T wave oversensing (3 patients). In the 15 patients, re-programming of the tachycardia detection interval and/or additional treatment with β-blocking agents were effective. In the 3 patients with T wave oversensing, the arrythmia was associated with an increase in T wave amplitude, change in T wave morphology and decreased R wave amplitude, and re-programming of the sensitivity of the local electrogram or changing the number of intervals to detect ventricular tachycardia decreased the number of inappropriate discharges in all 3 patients. In conclusion, inappropriate therapies are common problems in patients treated with the fourth generation ICD system, but most of them can be resolved using the dual-chamber ICD system. However, in patients with T-wave oversensing, it is difficult to avoid inappropriate discharge completely, even if the dual-chamber ICD system is implanted. (Jpn Circ J 2001; 65: 927 - 930)