Clinical Investigation
Inappropriate Discharges by Fourth Generation Implantable Cardioverter Defibrillators in Patients With Ventricular Arrhythmias
Takashi WashizukaMasaomi ChinushiMinoru TagawaHidehiro KasaiHiroshi WatanabeYukio HosakaFumio YamashitaHiroshi FurushimaAkira AbeHiroshi WatanabeJunichi HayashiYoshifusa Aizawa
Author information
JOURNAL FREE ACCESS

2001 Volume 65 Issue 11 Pages 927-930

Details
Abstract

The study prospectively investigated the incidence, cause and efficient management of inappropriate discharge by the fourth generation implantable cardioverter-defibrillator (ICD) system in 45 patients (mean age, 57±16 years). During the follow-up period of 27±17 months, 18 patients (40%) experienced one or more inappropriate therapies: sinus and supraventricular tachycardia (15 patients) and T wave oversensing (3 patients). In the 15 patients, re-programming of the tachycardia detection interval and/or additional treatment with β-blocking agents were effective. In the 3 patients with T wave oversensing, the arrythmia was associated with an increase in T wave amplitude, change in T wave morphology and decreased R wave amplitude, and re-programming of the sensitivity of the local electrogram or changing the number of intervals to detect ventricular tachycardia decreased the number of inappropriate discharges in all 3 patients. In conclusion, inappropriate therapies are common problems in patients treated with the fourth generation ICD system, but most of them can be resolved using the dual-chamber ICD system. However, in patients with T-wave oversensing, it is difficult to avoid inappropriate discharge completely, even if the dual-chamber ICD system is implanted. (Jpn Circ J 2001; 65: 927 - 930)

Content from these authors
© 2001 THE JAPANESE CIRCULATION SOCIETY
Next article
feedback
Top