原著
人工膝関節の術前プランとしての新しい大腿骨顆部撮影法
森実 和樹高橋 敏明竹田 治彦坪井 一世今井 浩山内 隆間島 直彦渡部 昌平山本 晴康
著者情報
ジャーナル 認証あり

2006 年 18 巻 1 号 p. 77-80

詳細
抄録
Precise femoral component setting is important for smooth patella tracking and good ligament balance in total knee arthroplasty. Most surgeons often use the posterior condylar line (PC line) and set the femoral component of the implant at 3° external rotation from the line. In contrast, some reports recommend using the transepicondylar axis (TEA) of the patient as the indicator. We report a new radiographical technique for assessment of TEA of the femur in total knee arthroplasty. The patient lies in the supine position, and flexes the knee as much as possible. The X-ray is applied to the knee at right angles to the skin. We measured the external rotation angle between the PC and clinical TEA lines, and the internal rotation angle between the clinical TEA and anterior condylar lines (AC line) in 34 patients (43 knees). The former angle was 5.6°±2.8° and the latter was -6.3°±2.9°. Both angles had almost the same values as the ones measured on CT scans and Kneeling View radiographs. Our new radiographical technique seems useful for assessment of the transepicondylar axis of the femur in total knee arthroplasty.
著者関連情報