原著
人工股関節全置換術術中のL字型脚長測定器の検討
藤井 洋佑松下 具敬青木 大輔廣岡 邦彦甲斐 信生
著者情報
ジャーナル 認証あり

2013 年 25 巻 1 号 p. 5-8

詳細
抄録
We designed a caliper positioned parallel to the femoral axis during total hip arthroplasty, and evaluated its accuracy in comparison with the straight caliper employed previously. At the same time, we evaluated the degree of femoral offset.
The correlation between the limb length measured intraoperatively and that measured by radiography was improved by use of the new caliper (p < 0.01, r = 0.83), but the correlation with femoral offset was not so high. Although measurement of the limb length parallel to the femoral axis is accurate, but not for that of the degree of femoral offset. To measure femoral offset correctly, it is necessary to reproduce the pre-and post-measurement abduction positions.
著者関連情報
前の記事 次の記事