Journal of Electrophoresis
Online ISSN : 1349-9408
Print ISSN : 1349-9394
ISSN-L : 1349-9394
Technical Note
Toward high performance western blotting using vacuum-driven system and laser scanner; comparison between two signal detection methods based on chemiluminescent and immunofluorescent imaging
Chisato HaradaHideji TajimaSetsuo HirohashiTadashi Kondo
著者情報
ジャーナル フリー

2009 年 53 巻 4 号 p. 63-66

詳細
抄録
A vacuum-driven antibody-reaction system enables high-throughput western blotting, where the antibodies pass through protein-transferred membrane and antibody-antigen reaction is completed within 30 min. With the vacuum-based device, two protein detection methods compatible with by laser scanning were compared; one with enhanced chemiluminescence reaction and the other using fluorescence-labeled antibody. Considering sensitivity, quantitative capability, reproducibility and labor intensiveness, we concluded that fluorescence-based detection is more suitable in combination with the vacuum-driven antibody-reaction system and a laser scanner.
著者関連情報
© 2009 by Japanese Electrophoresis Society