FIA研究懇談会会誌
Online ISSN : 2433-7374
Print ISSN : 0911-775X
Stepwise Injection Photometric Determination of Phenol in Air
Original Paper
Andrey BulatovMichael SolovievAnastasia PetrovaAleksey MoskvinLeonid Moskvin
著者情報
ジャーナル フリー

2010 年 27 巻 2 号 p. 158-

詳細
抄録
An automated method for the photometric determination of phenol in air has been developed. This method includes liquid-absorption isolation of phenol from a gas phase. The isolation proceeds directly in a reaction vessel by a hydraulic scheme of a stepwise injection analyzer with subsequent generation of an analytical form reacted with 4-aminoantipyrine. The determination concentration range is from 10 to 200 μg m-3 at the sample volume of 4.8 L and time of concentration of phenol is 6 min. Time for analysis is 10 min.
著者関連情報
© 2010 Japanese Association for Flow Injection Analysis