1981 年 16 巻 2 号 p. 135-139
A new method for evaluating weak leakage gamma-ray dose without background measurement was developed. A pulse height distribution obtained by a 3″φ×3″ NaI (Tl) scintillation spectrometer was converted to the incident gamma-ray energy spectrum by the response matrix method. Weak leakage gamma-ray dose rate with background was directly calculated from the energy spectrum. On the other background exposure rate was evaluated from natural activity concentrations deduced by the peak of 40K, 214Bi (U) and 208Tl (Th) in the energy spectrum. The results of measurement showed that this method was practically satisfiable technique.