固体飛跡検出器のトラック計測における重なりの補正方法
藤元 憲三松村 一博土居 雅広小林 定喜
著者情報
ジャーナル フリー

1990 年 25 巻 2 号 p. 129-133

詳細
抄録

Solid state nuclear track detectors (SSNTD) record radiation in the form of tracks. In the case of high track density, however, it is not always possible to distinguish each track separately. The track density might then be underestimated unless the loss of track number due to overlapping is compensated. An elaborated “erosion” or curve fitting process is applied usually, for the separation of the overlapping tracks, to automatic track counting systems. This paper shows a much simpler correction method which was developed by the analogy of the correction equation for the dead time of GM counters. From a set of about 10 data obtained from high track density detectors, the equation for SSNTD can be determined by a least square fitting. Once the equation is found, true track density could be derived easily without any help of complex image processing or calculation, such as the erosion or curve fitting. This method also provides the privilege of extending the upper bound of the detection limit.