保健物理
Online ISSN : 1884-7560
Print ISSN : 0367-6110
ISSN-L : 0367-6110
屋外ラドンおよびラドン娘核種濃度の連続測定
山崎 直飯田 孝夫下 道國池辺 幸正
著者情報
ジャーナル フリー

1995 年 30 巻 2 号 p. 149-154

詳細
抄録

The concentrations of outdoor radon and its progeny were measured every three hours. An electrostatic collection method for radon and three count techniques for radon progeny were adopted to measure the concentrations continuously. The measurements carried out from Feburary 1991 to June 1992. The equilibrium equivalent thoron concentrations were also measured with a radon progeny monitor. The concentrations of radon and its progeny showed the same diurnal variation of a maximum in the morning and minimum in the evening. An averaged equilibrium factor (F) was 0.76 over the whole period. The ratio of radioactivity concentration, 222Rn: 218Po: 214Pb: 214Bi, was 1.00: 0.80: 0.78: 0.72 in outdoor air. In comparison with outdoor radon progeny, the outdoor concentration of thoron progeny was about 1/40 and human exposure was about one-tenth.

著者関連情報
© 日本保健物理学会
前の記事 次の記事
feedback
Top