Journal of Arrhythmia
Online ISSN : 1883-2148
Print ISSN : 1880-4276
Chairs: Yousuke Ishii (Japan), Yukio Sekiguchi (Japan)
Incidence and Management for Trouble-Shooting Associated with the Sprint-Fidelis ICD Lead
Kenichi IijimaMasaomi ChinushiAkinori SatoYukio HosakaDaisuke IzumiHiroshi FurushimaHiroshi WatanabeMasahito SatoKatsuya EbeHiroshi ShimizuKazuyoshi TakahashiYoshifusa Aizawa
著者情報
キーワード: lead trouble, ICD
ジャーナル フリー

2011 年 27 巻 Supplement 号 p. OP53_2

詳細
抄録
Method: We investigated the outcomes of 136 patients with Sprint-Fidelis in 5 centers in Niigata Prefecture. Results Incidence: Lead troubles occurred in 9 patients (6.6%). The mean period before the trouble was 31 months (16–57 months). Implantation: Among these 9 patients, the lead was implanted using a cut-down technique in 4, an extra-thoracic approach in 4 and a subclavian approach in 1. Discovery: The lead troubles in 5 were discovered at regular clinics. The other 4 patients emergently visited because of the lead integrity alert (LIA) or inappropriate ICD discharges. In the 9 patients, the average sensing integrity counter was 338/day and mean F-F interval, of which was recognized as NST, was less than 200 ms. The LIA: The LIA had been programmed in 7, and 5 of the patients visited before inappropriate ICD discharges. However, the other 2 elderly patients couldn’t notice the alert sound. Extraction: Lead extraction was attempted in 6 and 5 leads were removable without any complications. Conclusions: The incidence of troubles with Sprint-Fidelis is high, and it occurs in the relatively late phase. The LIA is useful to detect the lead problems but its efficacy may be limited in elderly patients.
著者関連情報

この記事は最新の被引用情報を取得できません。

© 2011 Japanese Heart Rhythm Society
前の記事 次の記事
feedback
Top