Japanese Journal of Cardiovascular Surgery
Online ISSN : 1883-4108
Print ISSN : 0285-1474
ISSN-L : 0285-1474
A Case of Non-Anastomotic False Aneurysm of Late Fiber Deterioration in Dacron Graft
Akifusa HariyaAtsushi YamaguchiHideo AdachiSeiichiro MurataMasahiko OkadaTakashi Ino
Author information
JOURNAL FREE ACCESS

2001 Volume 30 Issue 2 Pages 95-98

Details
Abstract
Dacron prostheses are the most widely used grafts in replacement procedures for abdominal aortic aneurysms, but they are not perfect grafts. We encountered a rare case of late graft complication. A 66-year-old man was admitted to our hospital with a pulsatile mass in an abdominal operation scar. He had received placement of a Y-shaped Cooley double velour knitted Dacron graft 18 years previously. Computed tomography and angiography demonstrated graft dilatation and an aneurysm. After resection of the graft aneurysms, the operative findings showed a non-anastomotic aneurysm formation due to longitudinal division near the graft guideline. In this case, this graft failure may have been due to the deterioration of the filter of the Dacron prosthesis itself. Therefore it is important to perform careful long-term follow-up in patients with implanted Dacron arterial prostheses.
Content from these authors
© The Japanese Society for Cardiovascular Surgery
Previous article Next article
feedback