日本医学放射線学会物理部会誌
Online ISSN : 2186-9847
Print ISSN : 0288-5506
ISSN-L : 0288-5506
CaSO4:Tm+6LiおよびCaSO4:Tm+7Liを用いたTLD中性子線量当量計
都丸 禎三尾内 能夫入船 寅二内田 勲金田 浩一津屋 旭梅垣 洋一郎伊藤 彬
著者情報
ジャーナル フリー

1981 年 1 巻 1 号 p. 17-30

詳細
抄録

Response characteristics of two types of thermoluminescent neutron dosimeters were studied. One is a 30 cm-diameter polyethylene-paraffin (PEP) sphere with CaSO4: Tm+6Li (UD 136N) and CaSO4+7Li (UD 137N) at the center and the other is a 5cm-diameter PEP hemisphere containing six pairs of TLDs, in which four pairs are covered with cadmium and two pairs without cadmium. Comparison of sensitivities of CaSO4: Tm+6Li, Li2B407: Cu, and TLD 600 ribbon to thermal neutrons and gamma rays was also made. The 30 cm-sphere dosimeter was found capable of measuring 0.3 mrem of neutron dose in gamma background up to 1.2 mR. The 5 cm-hemisphere dosimeter was found suitable for neutron measurement received by personnel while handling 252Cf sources. The UD 136N was found to be 7 times more sensitive to thermal neutrons than TLD 600 ribbon. The results of measurements of practical neutron fields with these thermoluminescent neutron dosimeters were described. (This work was supported in part by the U. S. Department of Energy under contract No. EY-76-C-09-0739.)

著者関連情報
© 一般社団法人 日本医学物理学会
前の記事 次の記事