日本医学放射線学会物理部会誌
Online ISSN : 2186-9847
Print ISSN : 0288-5506
ISSN-L : 0288-5506
JARP及びAAPMプロトコールに従って数種の円柱形電離箱により測定した高エネルギーX線の校正点吸収線量の比較
都丸 禎三入船 寅二内田 勲尾内 能夫
著者情報
ジャーナル フリー

1988 年 8 巻 2 号 p. 97-110

詳細
抄録

Absorbed dose conversion factors, Cλ, for several types of commercially available cylindrical chamber for photon beams with quality varying between 60Co γ rays and 20 MV X rays were calculated according to a new Japanese Association of Radiological Physicists (JARP) protocol. The new protocol C, takes account of differences between the compositions of the chamber wall and the buildup cap. For 10 MV X rays, the Cλ for NE 2581 chamber having a wall of A-150plastic and Lucentine build-up cap is different 1.2% from that for JARP chamber having a wall and buildup cap of acrylic plastics. In order to determine the consistency of the absorbed dose given according to the JARP protocol, absorbed doses were determined under the reference condition in a phantom irradiated by 60Coγ rays, and 4.2 and 8.5 MV X rays, using seven types of ionization chamber. The range of variation of the absorbed doses obtained was less than 1.1%for all investigated beam qualities. The ionization chamber readings were converted also to absorbed doses according to the American Association of Physicists in Medicine (AAPM) protocol. These absorbed doses showed good agrement, within 0.7% with absorbed doses obtained using the JARP protocols

著者関連情報
© 一般社団法人 日本医学物理学会
前の記事 次の記事
feedback
Top