X-ray Exposure Dose Control for X-ray CT System
Hiroshi Takagi
Author information
JOURNAL FREE ACCESS

2002 Volume 22 Issue 2 Pages 81-87

Details
Abstract

Shortening scan time of CT scanner system has been evolved and increase in number of CT examinations has also been remarkable. This has been resulted from global recognition of usefulness of the CT examination, contrary to this merit, however, it is important to recognize the risk of x-ray exposure dose. Japan Industry Association of Radiological Systems (JIRA) in which CT manufacturers join has issued the concrete countermeasure and guidance for reduction in x-ray exposure dose in response to the ICRP90 Recommendation.
Current CT scanner systems provide the data related to x-ray exposure dose such as CTDIW for setting CT scan parameters. To reduce x-ray exposure dose against infant patient, the scan parameters specified to infant patient (CT infant protocol) can be provided.
Exposure dose by x-ray CT can be measured by the measurement method corresponding to IEC-60601-2-44 and by using phantom. CTDI measurement is made by CTDI100 that measures in a range of 100mm for all slice thicknesses, and absorbed radiation dose is converted to that of air.
Dose profile is measured by using multiple thermoluminescence dosimeter (TDL) chips. CT exposure dose data including CTDI100 and Dose profile are well-defined, and Dose Information Guide conforming to IEC-60601-2-44 is provided to user for the purpose of reducing x-ray exposure dose.
Studies by low dose (2.5 3mA) simulation for the purpose of reducing x-ray exposure dose in screening CT examination of lung cancer and development of ROI scan to reduce x-ray exposure dose in puncture under CT fluoroscopy have also been conducted.