校正深および便宜的な最大深が高エネルギー電子線の線量測定に与える影響についての評価
羽根田 清文畠山 典子小山 矩
著者情報
ジャーナル フリー

2005 年 25 巻 2 号 p. 59-67

詳細
抄録
The Japan Society of Medical Physics (JSMP) has published a new dosimetry protocol “JSMP-01” for the calibration of radiotherapy beams. This protocol provides a new definition of the calibration point (depth) in order to obtain the absorbed dose at a reference point (Dr) by the calculation in high energy electron beams. This study evaluated the difference in the absorbed dose at the depth dose maximum (Ddmax) and Dr calculated from the absorbed dose at the calibration point. Further, the difference in the absorbed dose (Ddmax) at “measured maximum depth (dmax)” was evaluated using “temporary maximum depth (dmax). ” In the experiment at a depth interval of 0.1 g cm2, no difference was observed between (Ddmax) and (Dr). However, in the experiment at a depth interval of 0.3 g cm-2 the differences between (Ddmax) and (Dr) increased to 6.4%and 5.2% at 4 MeV and 6 MeV, respectively. Subsequently, at all energy levels the difference between (Ddmax) and (Dr) of all energy was more than 3% at a depth interval of 0.4 g cm-2. The differences between (Ddmax) and (Ddmax). were 2.68% and 4.50% at 6MeV and 9 MeV, respectively, for this depth interval.
著者関連情報
© 一般社団法人 日本医学物理学会