平行平板形電離箱(ルース型およびアドバンスト・マーカス型)の校正定数比の実験的評価
福村 明史遠藤 真広金井 達明竹下 美津恵坂間 誠伊藤 彬高橋 豊保科 正夫草野 陽介高瀬 英輔Jayantha Gunaratne
著者情報
ジャーナル フリー

2005 年 25 巻 2 号 p. 75-79

詳細
抄録
A Japanese code of practice for clinical dosimetry, titled “Standard Dosimetry of Absorbed Dose in External Beam Radiotherapy” was published by the Japan Society of Medical Physics (JSMP) in 2002. It mostly followed IAEA Technical Reports Series No.398, which was based on ND,W,i.e., the calibration factor in terms of absorbed dose to water for a dosimeter. The Japanese primary standard dosimetry laboratory, however, has not supplied ND,W but NX, as the calibration factor in terms of exposure. The unique feature of the Japanese code was provision of a data table of calculated conversion factors, ND,W / NX values, for many types of ionization chambers, excluding new plane-parallel ionization chambers. This paper describes the experimental evaluation of the conversion factors for the new plane-parallel ionization chambers, such as the Roos-type and Advanced Markus chambers. The obtained ND,W / NX values for PTW 34001, Wellhöfier PPC 40 and PTW 34045 were 37.96 ± 0.19,37.85 ± 0.36 and 37.90 ± 0.26 (Gy/C kg-1), respectively. They agreed with estimations based on Monte Carlo calculations.