最適化アルゴリズムがIMRT治療計画の強度分布に及ぼす影響
島田 真理中村 光宏宮部 結城山本 時裕手島 昭樹成田 雄一郎溝脇 尚志永田 靖平岡 真寛
著者情報
ジャーナル フリー

2006 年 26 巻 3 号 p. 97-107

詳細
抄録
In inverse planning of IMRT, optimum intensity maps are generated using an optimization algorithm. In this paper, impacts of two different optimization algorithms on the intensity map in IMRT treatment planning were evaluated. These were from the steepest descent (SD) and simulated annealing (SA) methods. The following five patterns were compared: [1] SD with calculation time of 5 min; [2] SD with the terminal criterion based on cost function; [3] SA with calculation time of 5 min; [4] SA with the terminal criterion; and [5] SA with the terminal criterion using a smoothing filter. Differences of D95%fo r the planning target volume, V70Gfyo r the rectum wall and the bladder wall were up to 0.5,1.8 and 3.2 %, respectively in all patterns. The dosimetric impact was negligible. In contrast, generated intensity maps were sensitive to the algorithms. Intensity maps generated by SA tended to have much fluctuation due to numerical artifacts compared to those generated by SD. The difference in the profile was over 7 % between the algorithms. The smoothing filter decreased the fluctuation in intensity maps of SA. In conclusion, it is important to understand impacts of optimization algorithms on the intensity map and the dose distribution.
著者関連情報
© 一般社団法人 日本医学物理学会