ガラスアンプルを使用したインフューザーポンプへの調製における外部フィルターの必要性
黒潟 貴子岡村 智文山﨑 博史小川 丈彦宮野 直之尾家 重治
著者情報
ジャーナル フリー

2012 年 38 巻 1 号 p. 56-60

詳細
抄録
For preparation using an infuser pump, the use of a filter is recommended on package inserts. However, there are no concrete data on the necessity of the outside filters. Therefore, we compared the number of particles in the preparation fluid in the infuser pump with and without using a filter. Particles with a diameter of 1.3-100 μm were counted using a light blockage particle counter KL-04 (Rion K.K., Tokyo, Japan). Without using a filter, the mean numbers of particles with diameters of 1.3-5, 5-10, and > 10 μm were 187.4 ± 76.6, 6 ± 3.3, and 1 ± 1, respectively. Using a filter, the mean numbers of particles with diameters of 1.3-5, 5-10, and > 10 μm were 12.4 ± 4.2, 1.2 ± 0.8, and 0.2 ± 0.4, showing decreases in each particle size range including significant decreases in particles with diameters of 1.3-5 and 5-10 μm. The preparation using the outside filter was necessary for removing contaminating particles.