携帯型持続注入ポンプ内の薬液の混合性と安全な薬液充填法の検討
瀬名波 宏昌名徳 倫明廣谷 芳彦下村 一徳中西 晶子小角 麻衣土井 彩愛川口 進一高橋 幸一
著者情報
ジャーナル フリー

2013 年 39 巻 1 号 p. 39-44

詳細
抄録

It has been reported that 5-fluorouracil (5-FU) and saline were not mixed homogeneously in some portable infusion pumps when saline was loaded first and then 5-FU was added.We evaluated whether the homogeneity of the mixture can be improved by modifying the loading method. The ratio between a glucose solution instead of 5-FU and saline and the order of their loading were changed, and differences in the glucose concentration were evaluated by labeling with ethyl hydroxybenzoate. While characteristic unevenness of the concentration was noted when glucose solution was loaded after saline, as has been reported, the unevenness was reduced by mixing the contents by inverting the pump after loading regardless of the mixing ratio. Therefore, in loading drugs with high biological toxicity such as 5-FU, the homogeneity of the mixture is considered to be improved, and the safety of the procedure to be improved by reducing the risk of exposure if the drug is loaded after physiologic saline and mixed by inversion.