1991 年 17 巻 5 号 p. 388-394
The clinical advantage of fluorescence polarization immunoassay for determination of cyclosporine (CYA) based on monoclonal antibody (M-FPIA) was compared with high-performance liquid chromatography (HPLC), radio immunoassay on monoclonal antibody (M-RIA), and fluorescence polarization immunoassay on polyclonal antibodies (P-FPIA).To clarify the evaluation of this method (M-FPIA), whole blood levels of CYA in 24 renal transplant recipients, who were treated by CYA therapy, were determined by each analytical method.
The coefficients of variation obtained by M-FPIA for the within-run and between-run precisions were 2.1 to 2.5% and 1.5 to 3.2%, respectively.There was a significantly good correlation between the data of HPLC and those of M-FPIA (r=0.988), M-RIA and M-FPIA (0.988), P-FPIA and M-FPIA (0.942), however, the value of CYA by P-FPIA was 2-fold of the value of the drug by M-FPIA.The values with M-FPIA were approximately the same as those with HPLC and M-RIA.These results suggest that M-FPIA be very useful for the emergent and routine monitoring of CYA on the clinical practice.