濁度変化を指標とした線溶酵素活性の測定
中川 輝昭北澤 英徳小林 史明佐藤 均足立 伊佐雄上野 雅晴堀越 勇
著者情報
ジャーナル フリー

1995 年 21 巻 6 号 p. 502-505

詳細
抄録
The fibrinolytic activities of commercially available urokinase were compared using a platelet aggregometer, which is capable of measuring the change in turbidity during the fibrinogen dissolution process.
The fibrinolytic activities were evaluated from the half-lives of the turbidity ratio at 660 nm using fibrin as a plasmin substrate. A clear linear relationship was evidenced between the half-life of the turbidity ratio and fibrinolytic activity, conforming that this method is superior to others with regard to accuracy, sensitivity, reproducibility, and quickness of measurement.
Furthermore, no difference was seen between the fibrinolytic activities of two commercially urokinase preparations. In conclusion, the platelet aggregometer was found to be very useful devise for evaluating the fibrinolytic activity of urokinase.
著者関連情報
© 一般社団法人 日本医療薬学会