截頭円錐柱型義歯維持装置の維持力に関する実験的研究
後藤 英伸堀田 浩史白石 和宏山辺 芳久藤井 弘之永野 清司
著者情報
ジャーナル フリー

1989 年 33 巻 4 号 p. 878-884

詳細
抄録

This study was made to clarify the retentive force of a denture retainer, i.e., a cone-type stud and sleeve attachment. In order to measure the wearing effect, an engaging and disengaging cycle test of 1, 000 times was performed. The effectsof taper angle (cone angle), sleeve height and thrusting force at engaging stud and sleeve on retentive force were also analyzed.
1. The cycle test had little effect on the retentive force (P∼0.41).
2. The retentive force became greater with decrease of taper angle (P<0.0001, Fig. 4).
3. The effect of the sleeve height on retentive force differed with taper angle (P>0.0001, Fig. 6).
4. The retentive force became greater with increase of thrusting force (P=0.0009), and this effect was greater as the taper angle decreased (P<0.0001, Fig. 5).
5. The contributions of the factors were approximately 38% intaper angle, 30% in thrusting force and 2% in sleeve height (Table 1).
6. Itcould be interpreted that we had better use the retainer with taper angle of 6 to 8 degrees, although the retainer with taper angle of 5 degrees could be applicable to the tooth on which the less occlusal force transmitted.