Japanese Journal of Radiation Safety Management
Online ISSN : 1884-9512
Print ISSN : 1347-1503
ISSN-L : 1347-1503
Technical report
Basic Study on the Properties of a Simplified Scintillation Survey Meter for γ–Rays
Rio ISOBEMiku FURUKAWASaya ONORyota SHINDOKeisuke YAMAMOTOYohei INABAMasatoshi SUZUKIKoichi CHIDA
Author information
JOURNAL FREE ACCESS

2023 Volume 22 Issue 2 Pages 72-82

Details
Abstract

 The purpose of this study was to evaluate the measurement accuracy of a simple γ–ray scintillation survey meter by conducting basic investigations (reproducibility, energy characteristics, influence of β–rays, etc.). We evaluated the performance of the simplified survey meter by using three types of scintillation survey meters for γ–rays and compared it with an energy-compensation type scintillation survey meter for γ–rays. As a result, it was found that the simplified survey meter may underestimate or overestimate the energy of γ–rays, or the energy of γ–rays to be measured may be out of range, depending on the conditions of use.

Content from these authors
© 2023 Japanese Society of Radiation Safety Managenent
Previous article Next article
feedback