The impact of self-shielded cyclotron operation on small-animal PET/CT equipment installed nearby, on the floor just above
Hitoshi KuboTamaki OtaniHideki OtsukaMasafumi Harada
著者情報
ジャーナル フリー

2014 年 61 巻 1.2 号 p. 46-52

詳細
抄録
Purpose: The purpose of this study was to evaluate the impact of a cyclotron on small-animal PET equipment installed directly above the cyclotron. Methods: The cyclotron equipment was HM-12, which has two targets, and the PET/CT equipment was Inveon. The equipment was installed in conformity to Japanese law and regulations. Before installation of the PET/CT equipment, the radiation dose, radio waves, and static and fluctuating magnetic fields were measured at the position where it would be placed, both when the cyclotron was in use and when it was not in use. After installation of the PET/CT, natural background and emission counts were measured at the same place under the same conditions. Results: An increase of radiation dose was observed when the target nearest the PET equipment was used. There were no distinct effects of radio waves or static and fluctuating magnetic fields. A significant increase of emission counts, approximately 300 cpm, was observed when the nearest target was used. Conclusions: Though radio waves and static and fluctuating magnetic fields generated by running cyclotron had no influence, a significant increase in emission count was observed. Careful attention should be paid to this influence when very low-radioactivity PET measurements are done. J. Med. Invest. 61: 46-52, February, 2014
著者関連情報
© 2014 by The University of Tokushima Faculty of Medicine
前の記事 次の記事
feedback
Top