Insertion of a Fogarty catheter through a slip joint section for neonatal and infantile one-lung ventilation : a report of two cases
Hiroki YonezawaRyosuke KawanishiHideto SasakiYohei SogabeKentaro HirotaYasuto HondaKatsuya Tanaka
Author information
JOURNAL FREE ACCESS

2021 Volume 68 Issue 1.2 Pages 209-212

Details
Abstract

Here, we report two cases involving a neonate and child in which a slip joint section was used to thread a Fogarty catheter into the endotracheal tube for one-lung ventilation (OLV). Both the neonate and infant required OLV, and were placed under general anesthesia. A Fogarty catheter was used for OLV. The Fogarty catheter was passed into the intraluminal side of the endotracheal tube through a slip joint section. OLV was maintained successfully without severe air leakage or Fogarty catheter displacement. The neonate had been intubated pre-operatively with a 3.5-mm inner diameter endotracheal tube, and we used that tube. These cases indicate that the technique can be applied to pre-operatively intubated patients and does not require surgeons to exchange endotracheal tubes. Use of the slip joint section technique facilitates Fogarty catheter fixation without additional dead space. J. Med. Invest. 68 : 209-212, February, 2021