Journal of Microwave Surgery
Online ISSN : 1882-210X
Print ISSN : 0917-7728
ISSN-L : 0917-7728
子宮内膜症
2.45GHzマイクロ波を用いるMicrowave Endometrial Ablationのための子宮内膜用電極の開発
金岡 靖平井 光三石河 修荻田 幸雄
著者情報
ジャーナル フリー

2001 年 19 巻 p. 51-54

詳細
抄録
A curved microwave applicator was specifically developed for microwave endometrial ablation for the uterus with distorted cavity. The applicator was transcervically introduced to a uterine cavity of a surgical specimen with an intramural myoma node and adenomyosis, and 11 sites on the endometrium were coagulated by microwaves. The endometrium and neighboring myometrium 7mm in depth were degenerated at a power of 70W with a duration of 50sec. The endometrium at the cornus was found fully destroyed. Temperature increased less than three Celsius degrees on the serosal surface of the uterus at the end of microwave coagulation procedure. It is highly feasible to use the curved applicator for uteri enlarged with intramural or submucosal myomas in clinical setting.
著者関連情報
© 2001 特定非営利活動法人 Microwave Surgery研究会
前の記事 次の記事
feedback