Displacement and Stress Analysis around the Artificial Acetabular Cup in a Total Hip Replacement
Satoshi KAKUNAIHiroyuki TACHIBANATohru SAKAMOTOMasayoshi ABODaisaku IKEDAHiroo FUJIWARA
著者情報
ジャーナル フリー

2007 年 1 巻 3 号 p. 332-342

詳細
抄録
In order to improve the service life of the artificial acetabular cup in a total hip replacement, it is important to determine the best material and design, and to assess the mechanical behavior around the cup. In this study, electronic speckle interferometry (ESPI) and the two-dimensional finite element method (FEM) are employed to investigate the mechanical behavior. The influence of the cancellous bone and cup thickness on mechanical behavior around the cup was investigated. Good agreement of the cup model was found between the ESPI measurements and FEM predictions. The following results were obtained. (1) Cancellous bone with a porous structure can be measured by the ESPI method. (2) There are discontinuities of the displacement distribution in the transverse direction in each boundary region of the cup, bone cement and cancellous bone. (3) The maximum shear stress exists in the boundary region of the cup and bone cement.
著者関連情報
© 2007 by The Japan Society of Mechanical Engineers