日本応用磁気学会誌
Online ISSN : 1880-4004
Print ISSN : 0285-0192
ISSN-L : 0285-0192
生体磁気・磁場効果
脳腫瘍細胞刺激用パルス磁界発生装置の試作
内川 義則小林 宏一郎内田 靖浩杉山 聡宮下 收
著者情報
ジャーナル オープンアクセス

1997 年 21 巻 4_2 号 p. 737-740

詳細
抄録
This paper focuses of the trial production of a pulsing magnetic field generator for magnetic stimulations of organic tissues such as brain tumor cells (RGC-6). It consists of a combination of Helmholtz coils (radius: 20 cm, height: 27.5 cm, distance between coils: 6.5 cm) to generate the magnetic field, and a flip-flop-type electric circuit to generate the pulsing electric current controlled by switching of thyristors. This generator can supply a pulsing magnetic field of 10 mT at 200 Hz with a rise-time of 0.4 ms. The uniformity of the magnetic field distribution was within a maximum error of 5.8% in the 2400 cm3 region (20×20×6cm) between the Helmholtz coils.