Journal of Neuroendovascular Therapy
Online ISSN : 2186-2494
Print ISSN : 1882-4072
ISSN-L : 1882-4072
Original Articles
Review of Distal Embolism with Respect to Filters for Carotid Artery Stenting
Kentaro HayashiYuki MatsunagaYukishige HayashiKiyoshi ShirakawaMitsuto Iwanaga
著者情報
ジャーナル オープンアクセス

2019 年 13 巻 1 号 p. 9-15

詳細
抄録

Objective: Carotid artery stenting is performed using a device for preventing distal embolism because vasodilation-related debris may cause cerebral infarction. Concerning filters for preventing embolism, membrane-type filters have been used, but mesh-type filters became commercially available. We have selected filter-assisted stenting as a first-choice procedure. We examined post-treatment filters under a microscope, and reviewed the pathogenesis of distal embolism.

Methods: The subjects were 83 patients in whom carotid artery stenting with a filter was performed, and filters could be examined after surgery (Angioguard XP [AG; Cordis Corporation, Miami Lakes, FL, USA]: 25 patients, Filterwire EZ [FW; Boston Scientific, Natick MA, USA]: 32, and Spider FX [Spider; Covidien, Dublin, Ireland]: 26). After treatment, the filters were stained with hematoxylin and eosin (HE), separated from the struts, and embedded in preparations for microscopic observation. Debris was classified into plaque-derived and fibrin-formation types, and quantified as an area using computer software. Distal embolism was evaluated based on intraoperative flow impairment, postoperative symptoms, and perioperative diagnostic imaging findings.

Results: Intraoperative flow impairment was noted in six patients (24%) in the AG group, five (15.6%) in the FW group, and one (3.8%) in the Spider group. Cerebral infarction was observed in three (12%), two (6.3%), and two (7.6%) patients, respectively. There were no differences in the volume of plaque-derived debris, but the volume of fibrin-formation-type debris was more in the AG group. As a result, the volume of debris collected was more. In the Spider group, the volume of fibrin-formation-type debris was minimum.

Conclusion: Functions differed between the membrane-type and mesh-type filters. Considering their performance, these filters should be used.

著者関連情報
© 2019 The Japanese Society for Neuroendovascular Therapy

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
前の記事 次の記事
feedback
Top