Journal of Nuclear Science and Technology
Online ISSN : 1881-1248
Print ISSN : 0022-3131
Gamma-Ray Absorbed Dose Measurements in Mediawith Plural Thermoluminescent DosimetersHaving Different Atomic Numbers
Shun-ichi TANAKANobuo SASAMOTO
Author information
JOURNAL FREE ACCESS

1985 Volume 22 Issue 2 Pages 109-119

Details
Abstract

Corrective factors for thermoluminescent dosimeters to obtain the γ-ray dose in materials are examined systematically based on a general cavity ionization theory, and an experimental method; interpolation method, is presented for measuring the γ-ray absorbed dose in materials using thermoluminescent dosimeters with different atomic numbers. A characteristic feature of the method is to measure the γ-ray absorbed dose in materials without any γ-ray spectrum information. In order to examine the feasibility, the absorbed doses are measured in iron, lead and polyethylene media with a 60Co γ-ray source, and compared with a transport cal-culations whose accuracy is already verified. It is shown that the γ-ray absorbed doses meas-ured using the present method agree with the calculated ones within ±5 to ±10%.

Content from these authors

This article cannot obtain the latest cited-by information.

© the Atomic Energy Society of Japan
Previous article Next article
feedback
Top