Application of Tail Subtraction Technique with Particle Identification to a Stacked Spectrometer for Intermediate Energy Protons
Hiroki MUROHKAHiroki YOSHIDAYusuke UOZUMIAkihiro NOHTOMITakeji SAKAEMasaru MATOBAOsamu IWAMOTOTakashi MAKINorihiko KOORI
著者情報
ジャーナル フリー

1997 年 34 巻 7 号 p. 708-713

詳細
抄録

A solid detector for intermediate energy protons needs a thick material to stop incident protons. The response function of detector has a full-energy peak and a tail when monochromatic energy protons are measured by the detector. A tail subtraction technique with particle identification was applied to remove the tail. The Gaussian distribution was fitted to a peak in a particle identification spectrum. In this technique, the window of three times of standard deviation in the peak region was found to be proper in anayzing the particle identification spectrum. Using this technique, the tail could be removed. The peak efficiencies of the stacked spectrometer for protons were obtained up to 250MeV.