Journal of Japan Oil Chemists' Society
Online ISSN : 1884-1996
Print ISSN : 1341-8327
ISSN-L : 1341-8327
Determination of Lipase Activity by High Performance Liquid Chromatography
Makoto FUKATSUChales T. THORSTEINSONJames K. DAUNToshitake TAMURA
Author information
JOURNAL FREE ACCESS

1997 Volume 46 Issue 6 Pages 697-700,715

Details
Abstract
Lipase unit is defined as that amount of enzyme needed to produce 1 μmol non-esterified fatty acid in one minute under specified conditions. Non-esterified fatty acid is traditionally measured by acid-base titration. This was done in the present study in place of High-performance liquid chromatography (HPLC) method. HPLC with a light scattering detector was conducted to measure non-esterified fatty acids to determine the activity of lipase from Candida cylindracea. HPLC result was essentially the same as that obtained by titration, the former method being more sensitive.
Content from these authors