Journal of Oral Science
Online ISSN : 1880-4926
Print ISSN : 1343-4934
ISSN-L : 1343-4934
Original Article
Utilization of a mask fitter or micropore tape to improve the fit of a surgical mask
Kwanchanok YoucharoenPatchara TaosiriPruettinun MuangthongAtiya PoenateetaiNonglak SombunthamNirada Dhanesuan
Author information
JOURNAL FREE ACCESS

2022 Volume 64 Issue 4 Pages 271-273

Details
Abstract

Purpose: This study aimed to test the face seal of a surgical mask modified using a custom-made mask fitter or by sealing all borders with micropore tape, in comparison to the N95 mask as a gold standard.

Methods: Fifteen participants were assigned to wear an N95 mask, a surgical mask sealed with a mask fitter, and a surgical mask sealed with micropore tape. A quantitative fit test was performed using a Portacount respirator fit tester in 4 different actions: bending over, talking, moving the head from side to side, and moving the head up and down.

Results: The N95 showed the highest overall fit factor score (134.67 ± 66.62), passing Occupational Safety and Health Administration (OSHA) standards. The surgical mask alone had the lowest overall fit factor score of 4.73 ± 3.30. Modification of the surgical mask using a mask fitter or micropore tape significantly increased the overall fit factor to 35.33 ± 14.58 and 29.33 ± 9.73, respectively. This pattern was similar for all exercises.

Conclusion: The N95 was the only mask type that passed the OSHA standard for protection. Use of a mask fitter or micropore tape significantly increased the face seal of a surgical mask. This could offer useful levels of protection during a non-aerosol dental procedure.

Content from these authors
© 2022 by Nihon University School of Dentistry
Previous article Next article
feedback
Top