Development of an Automated Sample Injection System for Pillar Array Columns
Hiroshi KUROKIHirotaka KOYAMAYusuke NAKATANITakashi FUNATSUShigeyoshi HORIIKEMakoto TSUNODA
著者情報
ジャーナル フリー 早期公開

論文ID: 2019.022

この記事には本公開記事があります。
詳細
抄録

A pillar array column with a perfectly ordered internal structure has a higher separation efficiency than a particle-packed column used in conventional high-performance liquid chromatography. However, applying the pillar array column technology to quantitative analysis is challenging as the injection volume of the sample varies for each injection. This occures because the sample volumes are in the order of nanoliters and the sample is injected manually. In this study, an automated sample injection system was developed to solve this problem. The system was composed of two pumps (one for the sample and other for the mobile phase), a six-way valve that can be controlled by a PC, and an autosampler. The peak height deviation, which is more than 20% in the conventional manual injection system, was improved to 1.2% and 0.4% for the two coumarin dye samples. This result indicated that the automated sample injection method developed in this study could be applied to pillar array columns to allow for quantitative analysis.

著者関連情報
© 2019 The Society for Chromatographic Sciences