Effect of tightening protocols on abutment screw loosening and morphological changes in two-piece zirconia implants: A comparative study using cyclic torsional loading
Toma Hara , Takayuki Sugiki, Munetsugu Seto, Kazuhiko Ueda
Author information
JOURNAL OPEN ACCESS

2026 Volume 70 Issue 3 Pages 468-477

Details
Abstract

Purpose: This study evaluates the effects of different abutment screw tightening protocols and implant materials on screw loosening and morphological changes after cyclic torsional loading of two-piece implants.

Methods: Titanium (T) and zirconia (Z) two-piece implants were used. Abutment screws were tightened to 35 Ncm. Three protocols were tested: (A) retightening after 5 min, (B) immediate retightening, and (C) no retightening (n = 6 each). All samples underwent 100,000 cycles of cyclic torsional loading. Removal torque values before and after loading (RT-pre and RT-post) and torque loss percentage (TL) were calculated. Surface morphology was assessed using scanning electron microscopy (SEM), and dimensional changes were quantified using root mean square (RMS) values from three-dimensional data.

Results: In the titanium group, protocol A (TA) had lower RT-post values than protocols B (TB) and C (TC) (P < 0.001), and TB was lower than TC (P < 0.05). In the zirconia group, protocol A (ZA) was significantly lower than protocols B (ZB) and C (ZC) (P < 0.05, P < 0.001, respectively). RT-post and RT-pre were significantly lower in the zirconia group than in the titanium group under all conditions (P < 0.001). No differences were observed in TL. SEM revealed wear on the screw heads and threads in all groups. RMS values were significantly lower in the zirconia group (P < 0.05).

Conclusions: Abutment screw retightening does not consistently enhance torque maintenance under torsional loading. The implant material affects torque retention and deformation, where titanium implants demonstrating greater stability than zirconia implants.

Content from these authors
© 2026 Japan Prosthodontic Society

This is an open-access article distributed under the terms of Creative Commons Attribution-NonCommercial License 4.0 (CC BYNC 4.0), which allows users to distribute and copy the material in any format as long as credit is given to the Japan Prosthodontic Society. It should be noted however, that the material cannot be used for commercial purposes.