Journal of Physical Therapy Science
Online ISSN : 2187-5626
Print ISSN : 0915-5287
ISSN-L : 0915-5287
Original Article
Changes in cervical muscle activity according to the traction force of an air-inflatable neck traction device
Jong Ho KangTae-Sung Park
著者情報
ジャーナル フリー

2015 年 27 巻 9 号 p. 2723-2725

詳細
抄録

[Purpose] The purpose of this study was to analyze cervical muscle activity at different traction forces of an air-inflatable neck traction device. [Subjects] Eighteen males participated in this study. [Methods] The subjects put on an air-inflatable neck traction device and the traction forces administered were 40, 80, and 120 mmHg. The electromyography (EMG) signals of the splenius capitis, and upper trapezius were measured to assess the muscle activity. [Results] The muscle activity of the splenius capitis was significantly higher at 80, and 120 mmHg compared to 40 mmHg. The muscle activity of the upper trapezius did not show significant differences among the traction forces. [Conclusion] Our research result showed that the air-inflatable home neck traction device did not meet the condition of muscle relaxation.

著者関連情報
© 2015 by the Society of Physical Therapy Science. Published by IPEC Inc.

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (by-nc-nd) License.
https://creativecommons.org/licenses/by-nc-nd/4.0/
前の記事 次の記事
feedback
Top