Physical and Histopathological Assessment of the Effects of Metallic Stents on Radiation Therapy
Yoshihiko TSUJIHitoshi YOSHIMURAFumiaki UTOToshiaki TAMADAKazuro IWATATetsuro TAMAMOTOIsao ASAKAWATakayuki SHINKAIKimihiko KICHIKAWAMasatoshi HASEGAWA
著者情報
ジャーナル フリー

2007 年 48 巻 6 号 p. 477-483

詳細
抄録
To evaluate whether simultaneous metallic stent (MS) placement and radiotherapy are feasible, phantom and animal experiments were performed. The interface dose by external irradiation (EI) or intracavity irradiation (II) to 5 kinds of MS was measured using the charge-coupled device (CCD) camera with a thermoluminescent (TL) sheet, and backscatter and absorption were evaluated using composite method. Lineac 10 MV X-ray irradiated the MS in close contact with the TL sheet. II was performed using 192Ir, and the irradiation dose transmitted through the MS was measured using the TL sheet. The ratio of the CCD value of the MS wire region to that of the MS non-wire region was defined as the dose perturbation factor (DPF). Furthermore, the effects of a combination of 60Co gamma-ray EI and MS placement in the normal common bile duct were histopathologically evaluated in dogs. In the phantom experiments of EI, in backscatter by the MS, the DPF was 1.09 for CZ, and 1.03 for Pal, but no backscatter was detected in the remaining 3 MS. In absorption by the MS, the DPF was 0.92, 0.97, 0.97, and 0.98 for CZ, Wall, Pal, and Vel, respectively, but no absorption was detected in U. Flex. In those of II, the DPF of absorption was 0.91, 0.98, and 0.98 for CZ, U. Flex, and Wall, respectively, but no absorption was detected in Pal and Vel. The animal experiments showed infiltration of inflammatory cells and fibrosis in the case of both MS placement and EI. These changes were marked in EI treating after MS placement, but neither severe ulcer nor perforation was found. In conclusion, these results suggested that the effect of MS should be considered carefully when simultaneous MS placement and EI is performed clinically.
著者関連情報

この記事は最新の被引用情報を取得できません。

© 2007 by Journal of Radiation Research Editorial Committee
前の記事 次の記事
feedback
Top