Journal of the Robotics Society of Japan
Online ISSN : 1884-7145
Print ISSN : 0289-1824
ISSN-L : 0289-1824
Paper
Development of Master Slave System for Interventional Radiology with Force-Rate Control
Masaru IdePierluigi Beomonte ZobelPaolo Di ClaudioMakoto MohriTakashi Komeda
Author information
JOURNAL FREE ACCESS

2010 Volume 28 Issue 2 Pages 223-230

Details
Abstract
The objective of this study is to develop a master-slave system for a catheter-guided operation, which is performed by using radiology, through the vascular system. When the master-slave system is used, the surgeon is not exposed to x-rays during the operation. The master tool is managed by an operator away from the slave tool, which is near the patient. The system must provide a realistic picture to the surgeon, particularly in term of force information because this operation is performed by observing three-dimensional fields on a two-dimensional monitor. In this paper, we describe the development of a master slave system that involves the use of force-rate control for guiding the catheter without using force sensors. The master tool has a force-display function. This system can be controlled by force and velocity controlling; hence, this system realized an innovative mechanism and algorism. Finally, the preliminary experiment indicated that the new control method was effective. Further, the force display was stable and achieved fast response.
Content from these authors
© 2010 The Robotics Society of Japan
Previous article Next article
feedback
Top