日本AEM学会誌
Online ISSN : 2187-9257
Print ISSN : 0919-4452
ISSN-L : 0919-4452
学術論文
体内深部-体外間の容量結合型無線情報伝送システム
-電磁ノイズ環境下における通信特性の測定-
新郷 航希柴 建次原 一裕坪木 光男伊藤 雅一
著者情報
ジャーナル フリー

2018 年 26 巻 1 号 p. 152-159

詳細
抄録

 Recently, many medical electronic devices have been developed that can be implanted deep inside the body. The devices are designed to transmit information collected from deep inside the body for receiving antennas outside the body. Therefore, we proposed a wireless digital information transmission system that transmits datum from deep inside the body to the outside using capacitive coupling. Inverting amplifier and low-pass filter (LPF) circuit in the receiving circuit are necessary for shaping distorted waveform before transferring it to the peripheral interface controller. To determine the optimum transmit filter constant (e.g., cutoff frequency and amplification rate), communication accuracy is measured using a prototype digital information transmission and receiving circuit and simulated human body. The optimum setting value of the cutoff frequency and the amplification rate were determined to be 17 kHz and 20 times, respectively. However, the communication accuracy reduced by approximately 40% when electromagnetic noise voltage (1.8 V, 50–100 kHz) was added to the receiving wave. To overcome this problem, a common-mode choke coil was inserted in front end of the receiving circuit. Thereby, communication accuracy was improved to approximately 98.7%.

著者関連情報
© 2018 日本AEM学会
前の記事 次の記事
feedback
Top