カプセル内視鏡用磁気誘導平面配列磁石の設計
松橋 華世小澤 悠平田代 晋久脇若 弘之水野 勉大宮 直木
著者情報
ジャーナル フリー

2021 年 29 巻 1 号 p. 13-18

詳細
抄録

 The purpose of this report is to design a magnetic guiding device for guiding capsule endoscopes. The goal is to obtain 17 mT at a distance of 100 mm from the guiding device. Furthermore, the device needs to be lightweight. Two magnetic guiding devices were proposed. The one with 4 magnets arranged was 1475 g, and the one with donut type was 1229 g. In both cases, a magnetic flux density of 17.1 mT was obtained at a distance of 100 mm. However, the magnitude of the magnetic gradient was different, and the magnetic force obtained was 5 mN for the 4 cylindrical magnets and 5.9 mN for the donut type.