Animal Test and Clinical Application of Hollow Fiber Oxygenator for 100 cases of Open Heart Surgery
T. TSUJIK. SUMAY. TAKEUCHIY. INOUEK. SHIROMAT. YOSHIKAWAJ. NARUMIN. ITOH. KOBAYASHIH. FUKAZAWAH. HASEGAWAM. KANNOK. TANISHITA
Author information
JOURNAL FREE ACCESS

1981 Volume 10 Issue 1 Pages 167-169

Details
Abstract
The polypropylene hollow fiber oxygenator (H. F. O.) was tested in animal experiments and applied to 100 cases in open heart surgery.
8 dogs were used as test animals for H. F. O. using veno-arterial bypass method. The 1.0 M2 H. F. O. was perfused 5 hrs continuously with an average of 1.0 l/min of venous blood at 37°C and pure oxygen at a rate of 1.0 l/min. Average O2 transfer rate was 43±3ml/min/M2 and CO2 transfer rate, 43±2ml/min/M2 when Hb was 10.5g/dl. Free hemoglobin after 5 hr perfusion was 38.0±16.6 mg/dl. 100 cases including 77 congenital and 23 acquired heart diseases were operated using 1.8 or 3.3 M2 H. F. O.. Perfusion time was 74±66 min and free hemoglobin was 41.6±34.4mg/dl at the end of CPB. Average blood perfusion rate was 2500± 1000ml/min and pure oxygen gas to blood flow ratio was 0.53±0.21. PaO2 was 467±62 mmHg and PaCO2 29±5mmHg when average Ht was 24±5%.
This H. F. O. was observed to have excellent capability of oxygen and carbon dioxide gas transfer with small amount of hemolysis in 100 patients undergoing open heart surgery