人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
補助心臓非同期駆動時に見られた自然心心拍数の同期現象とその解析
高橋 英嗣佐久間 まこと山本 克之三田村 好矩三上 智久
著者情報
ジャーナル フリー

1981 年 10 巻 1 号 p. 27-30

詳細
抄録
Although the clinical application of LVAD has started, we have only a little knowledge on the hemodynamic changes during LVAD pumping. The interaction between the natural heart and LVAD was investigated through the entrainment of the heart rate by LVAD asynchronous pumping.
Sinus rhythm was monitored in LVAD implanted dogs and sheep while changing the pumping rate. The entrainment of sinus rhythm by LVAD pumping was observed in anesthetized dogs and also awake sheep. Namely, sinus rhythm varied according to the change in pumping rate when the pumping rate was approximately equal to or half the sinus rhythm.
To explain the discovered entrainment, computer simulation was performed. The macroscopic model of 'External periodic force and Self-oscillatory system' was used, since it has been known to induce the entrainment. In the model SA node was assumed to be a self-oscillatory system and LVAD an external periodic force. The SA node was also assumed to have the characteristic that sinus rhythm changes depending on the phase when neural impulses happen to arrive at the SA node. The computer simulation showed the similar modification of sinus rhythm by LVAD pumping to that in experiment.
The above gives us some idea of the control system of the heart and, moreover, LVAD is a useful tool for the physiological investigation.
著者関連情報
© 一般社団法人 日本人工臓器学会
前の記事 次の記事
feedback