人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
人工肝補助装置の研究
―アンモニア除去を中心として―
川西 秀樹西亀 正之児玉 求江崎 治夫土谷 太郎中光 晴彦
著者情報
ジャーナル フリー

1981 年 10 巻 2 号 p. 546-549

詳細
抄録
A new ammonia adsorption system was developed using silicate complex (TTY 500) and evaluated by means of in vitro and in vivo study. In vitro study evidenced that ammonia adsorption rate was 72.2%. In vivo study showed that serum ammonia in hepatic failure dogs decreased during the hepatic assist.
Direct hemoperfusion with charcoal coated by polyurethane (AAC) was carried out for the hepatic failure dogs. Satisfactory blood biocompatibility was obtained. The AAC-DHP group (4 dogs) at this prognosis was compared to the control group (17 dogs). During DHP there was significant decrease in serum tryptophan but no significant difference between both groups in survival time.
著者関連情報
© 一般社団法人 日本人工臓器学会