Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Development of a new model of Dynamic Patch Artificial Myocardium (Saitama-Toray model)
S. KYOH. EMOTOH. YAMANAKAS. TAKAMOTOY. YOKOTER. OMOTOY. MORI
Author information
JOURNAL FREE ACCESS

1984 Volume 13 Issue 1 Pages 174-177

Details
Abstract

We have already demonstrated that AVCO-MGH Dynamic Patch has beneficial effects for the recovery of the ischemic failing heart. However, its disadvantageous technical difficulty of patch anastomosis to LV must be improved for the clinical use of Dynamic Patch. Newly developed Saitama-Toray Dynamic Patch, the fragile dynamic membrane part is separated from the plastic frame for anastomosis, is designed to have regard especially for easiness and safety in patch anastomosis. Dogs experiment demonstrated that the hemodynamic effects of Saitama-Toray model is comparable to that of AVCO-MGH model in despite of its smaller volume displacement (25ml vs. 13ml).

Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top