Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Clinical evaluation of pulse generator for hemodialysis
T. MIYAZAKIH. NAITOS. OBATAKu. HARUNAY. OBATAKa. HARUNAK. FUKUNAGAY. KIRITAT. YAMANE
Author information
JOURNAL FREE ACCESS

1984 Volume 13 Issue 2 Pages 769-771

Details
Abstract
We clinically tried the “Pulse Generator” easily attached with ordinal roller pump for hemodialysis treatment. Comparison between the group of pulsatile flow and the group of no pulsatile flow, there are apparently differences in clearance and ultrafiltration. The “Pulse generator” can actively give pulsatile flow to get high clearance and high ultrafiltration using the same area of dialyzer, to aim at the short-time dialysis, or to promote high efficiency using for wearable artificial kidney.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback