Abstract
The most serious complication in CAPD today, is peritonitis. In order to decrease the frequency of peritonitis caused by accidents during the exchange of dialysis solutions, we have developed a new connecting device (FLAMELOK®) which makes use of flame sterilization. Clinical tests were performed on this device, and we have been able to achieve a decrease in the occurrence of peritonitis. However, fracture, crack and choke of the ceramic joints took place. Accordingly, the ceramic joints were improved. On confirmation of the durabiliby and safety of the new joints, we have observed an improvement in the heat resistance and intensity without a loss in the sterile effect.