Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Cineradiog raphic evaluation of the opening angle and ring motion of the Björk-Shiley valves in the aortic position
K. AOKIT. MIYAMOTOH. MURATAK. KAWAHARAY. OKAS. MAEDAK. YAMASHITAT. YOSHIDAK. KITAIH. TERAIM. TANAKAT. YASUOKAH. YAOH. SOH
Author information
JOURNAL FREE ACCESS

1985 Volume 14 Issue 3 Pages 1372-1375

Details
Abstract
The 22 Björk-Shiley valves were evaluated by cineradiographic examination. The opening angle was 58.6°±3.7° by Sohma's method and 59.3°±4.2° by Verdel's method. The ring motion was a 3°±1.5°. The 2 detached valves showed tilting motion of the ring and poor opening angle with fluttering phenomena of the disc.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top